Otometrics
69
Quality Tests
Electrode cable test
8.3
Electrode cable test
Every day before you start testing on patients, you should test the electrode cable to
make sure that it functions correctly.
2. If needed, press the
Start
button to
repeat the test.
3. Press the
OK
button to confirm the
test result.
1. Clip the electrode clips onto the metal bars of the ABR Tester.
2. Press the
Electrode cable
button on the
Quality Tests
screen.
3. Press the
Start
button to start the test.
If the electrode cables function correctly, the
message
Electrode cable OK
appears.