Tests
The TEOAE test
PASS (Test passed)
REFER (Test referred)
A
REFER (Test referred)
result may occur for a number of reasons, the most impor-
tant one being that the patient may have a sensorineural hearing loss.
Always consider the factors below, however, when you interpret a
REFER (Test
referred)
result. It may be possible to improve the quality of the test and so, on repeat-
ing the test a
PASS (Test passed)
may be obtained. It is always recommended that you
repeat the test when a
REFER (Test referred)
result is obtained.
Consider for instance the following when you interpret a
REFER (Test referred)
result:
• Malfunctions in the transmission of sound are by far the most important causes
of emissions not being detected. To eliminate this possibility, always make sure
If the result is a
PASS (Test passed)
, just the graph
is shown.
• A
PASS (Test passed)
result indicates that the
patient has normal outer hair cell function at
the time of testing.
• A
PASS (Test passed)
is determined by a statis-
tical algorithm, based on weighted averaging,
which ensures detection with proven high
specificity and sensitivity. No further inter-
pretation of a
PASS
result is needed.
If the result is a
REFER (Test referred)
, the
detailed results are shown immediately below the
graph.
REFER (Test referred)
:
• A
REFER (Test referred)
is “good” when the
Artifact
and
Stimulus
levels do not exceed
the critical limits (see below), indicating that
test conditions were good.
• A
REFER (Test referred)
is “bad” when the
Artifact
and
Stimulus
levels have been
exceeded, indicating that test conditions were
not optimal. In this case the patient should
be retested.
Содержание AccuScreen ABR
Страница 1: ...AccuScreen OAE ABR Screener User Manual Doc no 7 50 0920 US 02 Part no 7 50 09200 US 0459 ...
Страница 8: ...8 Otometrics Table of Contents AccuScreen OAE ABR Screener User Manual ...
Страница 66: ...66 Otometrics Printing Printouts from label printer ...
Страница 82: ...82 Otometrics Handling Data in AccuScreen Deleting data ...
Страница 88: ...88 Otometrics Maintenance and Cleaning Battery safety and maintenance ...
Страница 98: ...98 Otometrics App 2 Status and Error Messages Usage and test related messages ...