CHAPTER SEVEN
OAE test procedure
60
OAE test procedure
Consistent
Stimulus Out
Stimulus
Highlighted
High Noise
of Range
Ringing
Message
No
No
No
Checkfit
Yes No
No Noisy
No
Yes
No
Check Probe Fit
No
No
Yes Ringing
Yes
Yes
No
Check Probe Fit
Yes
No
Yes Ringing
No
Yes
Yes
Check Probe Fit
Yes
Yes
Yes
Check Probe Fit
The following table describes what
Highlighted Message
will appear if
more than one condition is met.
7.2
TE test (TEOAE model)
During testing the Otoport plays a series of clicks into the ear and records
the response. The OAE signal is very small and difficult to distinguish from
the noise made by other sounds in the room and by the patient. The longer
the Otoport records for, the easier it is to distinguish the OAE signal from
the noise. The Otoport looks at the ratio of the OAE signal to the noise;
this is known as the Signal to Noise Ratio or SNR. The Otoport measures
the SNR in a number of different frequency bands. The pass criteria is set
in the
Configuration
area (see chapter
11
Configuration
). The default
pass criteria is an SNR of 6dB and a minimum signal of >-5dBSPL in at
least two frequency bands.
It is the responsibility of the user to ensure that the pass criteria set meet
their requirements.
Содержание Otoport Lite
Страница 1: ...Otoport OAE ABR Otoport Lite model Issue 4 Otoport OAE ABR User Manual for Otoport Lite model...
Страница 4: ...CHAPTER ONE Introduction 4 Introduction...
Страница 6: ...CHAPTER ONE Introduction 6 Introduction...
Страница 130: ...CHAPTER TWELVE Printing and downloading 130 Printing 12 3 4 Sample prints Detailed print Summary print...
Страница 208: ...CHAPTER TWENTY FIVE Index 208 Index...