CHAPTER ELEVEN
Configuration
114
Configuration
The
minimum TC
requirement should be achieved when recording an ABR
waveform from a neonate (34 weeks to 6 months gestational age). The
strict TC
requirement is higher than largest TC ever likely to be achieved by
power line interference.
The two measures Fsp and TC are combined in four different ways as
follows:
PC1 requires:
Fsp level exceeds the dynamic Fsp with no template requirement
PC2 requires:
Fsp level exceeds the dynamic Fsp and TC exceeds the minimum TC
level
PC3 requires:
Fsp level exceeds the dynamic Fsp and TC exceeds the minimum TC
level
Or
Fsp level exceeds the minimum Fsp and TC exceeds the strict TC level
PC4 requires:
Fsp level exceeds the dynamic Fsp with no template requirement
Or
Fsp level exceeds the minimum Fsp and TC exceeds the strict TC level
When to use each PC
PC1 is ideal for performing an ABR test on adult subjects and is the
default setting for ABRcustom mode. It is useful for training, demonstration
and clinical investigations.
PC2 is designed for neonate screening and is the default setting for the
ABRscreen mode.
PC3 and PC4 may give higher specificity in challenging noise
environments, potentially passing healthy ears securely even when
interference levels are high. Both are designed for neonate screening but
PC4 can also be used with adult ears.
Содержание Otoport Lite
Страница 1: ...Otoport OAE ABR Otoport Lite model Issue 4 Otoport OAE ABR User Manual for Otoport Lite model...
Страница 4: ...CHAPTER ONE Introduction 4 Introduction...
Страница 6: ...CHAPTER ONE Introduction 6 Introduction...
Страница 130: ...CHAPTER TWELVE Printing and downloading 130 Printing 12 3 4 Sample prints Detailed print Summary print...
Страница 208: ...CHAPTER TWENTY FIVE Index 208 Index...