User Manual for Otoport Lite model
155
Probe care
16
Probe care
16.1
Cleaning
The following is the suggested method of cleaning an Otodynamics probe.
It should be noted that the probe is a precision assembly and, as such,
care should be taken throughout in its handling and cleaning.
Cable -
The cable may be cleaned with antiseptic fluid or wipes.
Probe casing -
The probe casing may be cleaned using antiseptic wipes
and dried with a tissue immediately afterwards. Do not allow liquids to
enter the sound tubes.
Coupler assembly -
Each coupler assembly has two sound tubes. These
are protected from ingress of foreign materials by wax guards in the tubes
and by the disposable probe tip. There is a loudspeaker at the end of one
tube and a microphone at the end of the other. Cleaning solution must not
penetrate the tubes.
16.2
Changing probe coupler tubes
The probe has sound tubes combined into a single coupler assembly that
can easily be replaced at regular intervals or when contaminated.
TRAINING
REQUIRED