Otoport OAE+ABR
User Manual for Otoport Lite model
105
Configuration
15 seconds of test time remaining. If the central soft key is pressed, an
E
will be displayed and the test will run until autostop criteria are met, it is
manually stopped or the absolute limit for testing is reached.
Start
With
Start
set to
On
(default), the testing will proceed from
Checkfit
to
Test
as soon as the Otoport detects a stable probe fit, without the need for
the user to select
Start
.
With
Start
set to
Off
, the user must initiate the test by selecting
Start
when
the probe fit and test conditions are satisfactory.
Irrespective of the
Start
setting, the
Start
key is available in order that the
user can initiate a test.