Mode of operation
23.2
DPOAE model
Parameter
Description
Checkfit stimulus
80µs positive broadband square wave pulse with an
intensity of 64dB pe (peak equivalent) in a 1cc cavity.
Sample rate
25.6kHz
Sample buffer
80ms (gives 2048 points)
FFT frequency bin
25Hz
DP Noise calculation
DP noise from the five spectral points above and
the five points below the DP frequency. In the
‘Rapid’ noise mode, the mean of these points gives the
noise level. In the ‘Standard’ noise mode the mean of
these points plus two standard deviations gives the
noise level.
Noise rejection calculation
The noise level for noise reject is calculated from
the difference between consecutive 80ms samples.
Stimulus frequencies
(F2)
Sine waves of 6, 4, 3 and 2 kHz
Frequency ratio
F2 = 1.22 F1
Memory capacity
Patients
500
Tests per patient
256
Total tests
500