3.5.3
Re-Calibrating Using the Calibration Mode
The
CALIBRATION MODE
allows you to recalibrate the instrument. First, perform the visual inspection of
the calibration state as explained above. If you experience a deviation between the target peak position
and the actual position, you can perform a manual recalibration. If “operative” is shown (Figure
, p.
)
a recalibration is not necessary. Otherwise, when “System can be improved” shows up, a recalibration is
necessary.
Figure 25:
Calibration check screen for
40
K
After entering the
CALIBRATION MODE
, it takes some time until the peak fit is established. The shown
percentage value represents the progress of acquiring the minimum counts to establish the measured
peak position. This depends on the strength of the calibration source you are using. Once the peak fit
quality is sufficient and enough counts are contained in the spectrum, the
CAL
button becomes active.
You can now press
CAL
to perform the manual recalibration and to definitively update the internal gain.
In the example in Figure
, p.
, we used
137
Cs. Calibrations can also be done on
40
K and
60
Co.
After recalibration, the calibration check acquisition is reset and you will get an updated view of the
peak fit. You can now again inspect the results of the recalibration.
3.6 The System Settings Screen
3.6.1 Setting Date and Time
Once
SYSTEM
is highlighted in the
ADVANCED MODE
, press
SELECT
to enter the
SYSTEM SETTINGS
dialogue. The button
NEXT
navigates through each user editable field of these settings.
Page 34 of 60
RADEAGLET User Manual• Software 2.1.16 • Document 1.32o • 2017-12-15
© innoRIID GmbH