digiBASE-E ™ High-Performance Power-over-Ethernet Multichannel Analyzer / PMT Base for Scintillation Detectors
54
SHOW_LIVE_PRES
Reports the current live-time preset in units of 20 ms (50 ticks/s). See also
CLEAR_PRESETS and SET_LIVE_PRESET.
Responses:
$G0000000000075<CR>
Live-time preset reported as disabled.
$G0000000001076<CR>
Live-time preset reported as 1 tick.
...
...
$G4294967295132<CR>
Live-time preset reported as 4294967295 ticks.
SHOW_LIVE_REMAINING
Reports the current live-time remaining in units of 20 ms (50 ticks/s). See also
SET_LIVE_REMAINING.
Responses:
$G0000000000075<CR>
Live-time remaining reported as disabled.
$G0000000001076<CR>
Live-time remaining reported as 1 tick.
... ...
$G4294967295132<CR>
Live-time remaining reported as 4294967295 ticks.
SHOW_LLD
Shows the lower level discriminator setting. See also SET_LLD.
Responses:
$C00050092
The LLD is 50.
SHOW_MODE
Reports mode of operation (List, PHA, or Sample [InSight]).
Response:
$FPHA<CR>
PHA mode.
$FLIS<CR>
List mode.
$FSAM<CR>
Sample (InSight) mode.
SHOW_MONI_LABEL num
Returns with an ASCII string label designation for the state-of-health parameter for num.
SHOW_MONI_MAX
Returns with the number of monitors available for viewing.
SHOW_MONI_VALUE num
Returns with an ASCII representation of the value for the monitor. For the digiBASE-E,
SHOW_MONI_VALUE 0 returns the 32-bit event counter value.