5. FIRMWARE COMMANDS AND RESPONSES
53
Bit 0 (LSB):
Bias supply polarity (0=positive, 1=negative).
Bit 1:
Bias supply overload (0=overload, 1=normal).
Bit 2:
HV enabled (0=disabled, 1=enabled).
Example Responses:
$D0100000003077<CR>
1000 V, negative, not overloaded, disabled.
$D0100000002076<CR>
1000 V, positive, not overloaded, disabled.
$D0100000007082<CR>
1000 V, negative, not overloaded, enabled.
SHOW_INTEGRAL [start_chan,number_of_chans]
Reports the sum of the specified group of spectral data channels. If start_chan and
number_of_chans is not provided, SHOW_INTEGRAL reports the sum of all channels in the
currently selected segment that have their ROI flag set.
Responses:
$G0000000000075<CR>
Integral reported as 0.
...
...
$G4294967294131<CR>
Integral reported as 4294967294.
$G4294967295132<CR>
Integral reported as greater than or equal to 4294967295
(maximum reportable value).
SHOW_INTEGRAL_PRESET
Reports the current ROI integral preset value. For more information about the ROI integral
preset, see SET_INTEGRAL_PRESET. See also SHOW_INTEGRAL.
Responses:
$G0000000000075<CR>
Integral preset reported as 0.
...
...
$G4294967295132<CR>
Integral reported as 4294967295.
SHOW_LENGTH_SAMPLE
Reports the number of points in the InSight Mode Waveform.
Response:
$C00256100
256 points in the waveform.
SHOW_LIVE
Reports the contents of the live-time counter in units of 20 ms (50 ticks/s). See also
CLEAR_COUNTERS and SET_LIVE.
Responses:
$G0000000000075<CR>
Live time reported as 0 ticks.
$G0000000001076<CR>
Live time reported as 1 tick (20 ms).
...
...
$G4294967295132<CR>
Live time reported as 4294967295 ticks (over 23000 days).