Operation
79
-
5.2 Aiming laser
The CTlaser has an integrated double laser aiming which helps for the alignment of the sensor. Within the
two laser points lies the measurement spot. At the focus point of the according optics
both lasers are crossing and showing as one dot the minimum spot. This enables an
alignment of the sensor to the object.
Do not directly point the laser at the eyes of persons or animals! Do not stare into the laser
beam. Avoid indirect exposure via reflective surfaces!
The two laser points mark the position of the measuring spot, but not its exact size. The exact
size of the measurement spot can be found in the optical charts
. As an
alternative to the optical diagrams, the spot size calculator can also be used on the optris
website
http://www.optris.com/spot-size-calculator
At ambient temperatures >50 °C the laser will be switched off automatically.
The laser should only be used for sighting and positioning of the sensor. A permanent use of
the laser can reduce the lifetime of the laser diodes.
Furthermore, in a permanent use of the laser, the measurement accuracy can be affected.