18
-
Shock
IEC 60068-2-27 (25G and 50G)
Software (optional)
CompactConnect
1)
Laser will turn off automatically at ambient temperatures >50 °C. The functionality of the LCD display can be limited at ambient
temperatures below 0 °C.
2.3 Electrical specifications
Power supply
8–36 VDC
Current draw
Max. 160 mA
Aiming laser
635 nm, 1 mW, On/ Off via programming keys or software
Outputs/ analog
Channel 1
selectable: 0/ 4–20 mA, 0–5/ 10 V, thermocouple (J or K) or alarm output (Signal source:
object temperature)
Channel 2 (LT/ LTF/ MT/ F2/ F6/ G5/ P7 only)
Head temperature [-20...180 °C] as 0–5 V or 0–10 V output or alarm output (Signal source
switchable to object temperature or electronic box temperature if used as alarm output)
Alarm output
Open collector (NPN type) output at Pin AL2 [24 V/ 50 mA]
Output impedances
mA
Max. loop resistance 500 Ω (at 8-36 VDC)