20
-
2.4 Measurement specifications [LT models]
LT
LTF
Temperature range (scalable)
-50...975 °C
Spectral range
8...14 µm
Optical resolution
75:1
50:1
System accuracy
1), 2)
(at ambient temp. 23 ±5 °C)
±1 °C or ±1 %
±1,5 °C or ±1,5 %
Repeatability
1)
(at ambient temp. 23 ±5 °C)
±0.5 °C or ±0.5 %
±1 °C or ±1 %
Temperature resolution (NETD)
0.1 K
0.5 K
Response time (90 % signal)
120 ms
9 ms
Warm-up time
10 min
Emissivity/ Gain, Transmissivity
0.100...1.100 (adjustable via programming keys or software)
Signal processing
Average, peak hold, valley hold (adjustable via programming keys or software)
1)
Accuracy for thermocouple output: ±2.5 °C or ±1
%,
whichever is greater
2)
at object temperatures >0°C, Ɛ =1