74
-
S ON
Activating (
ON
) and Deactivating (
OFF
) of the
Sighting Laser
. By pressing
Up
or
Down
the laser can be switched on and off.
ð MV5
Selection of the
Output signal
. By pressing
Up
or
Down
the different output signals can
E0.970
Setup of
Emissivity
. Pressing
Up
increases the value,
Down
decreases the value (also
valid for all further functions). The emissivity is a material constant factor to describe the
ability of the body to emit infrared energy
.
T1.000
Setup of
Transmissivity
. This function is used if an optical component (protective
window, additional optics e.g.) is mounted between sensor and object. The standard
setting is 1.000 = 100 % (if no protective window etc. is used).
A 0.2
Setup of
Average time
. In this mode an arithmetic algorithm will be performed to
smoothen the signal. The set time is the time constant. This function can be combined
with all other post processing functions. The shortest value is 0.001 s and can be
increased/ decreased only by values of the power series of 2 (0.002, 0.004, 0.008,
0.016, 0.032, ...). If the value is set to
0.0
the display will show
---
(function deactivated).
P----
Setup of
Peak hold
. If the value is set to
0.0
the display will show
---
(function
deactivated). In this mode the sensor is waiting for descending signals. If the signal
descends the algorithm maintains the previous signal peak for the specified time.
Содержание CTlaser 05M
Страница 1: ...Operator s Manual optris CTlaser LT LTF 05M 1M 2M 3M MT F2 F6 G5 P7 Infrared thermometer...
Страница 8: ...8 Appendix C Smart Averaging 96 Appendix D Declaration of Conformity 97...
Страница 44: ...44 Figure 2 Mounting bracket adjustable in one axis Order No ACCTLFB standard scope of supply...
Страница 45: ...Mechanical Installation 45 Figure 3 Electronic box...
Страница 97: ...Appendix D Declaration of Conformity 97 Appendix D Declaration of Conformity...
Страница 98: ...optris CTlaser E2018 12 A...