Keratron
™
Onda Installation and operating manual
OPTIKON 2000
Cod. 161401EN
5-7
2015-03-27
Rev.C
5.2
TECHNICAL SPECIFICATIONS
PARAMETER
SPECIFICATIONS
Manufacturer: .............................
OPTIKON 2000 S.p.A.
Via del Casale di Settebagni, 13
00138 Rome - Italy
Model: .........................................
Keratron
™
Onda
Regulatory conformity: ...............
Directive on medical devices 93/42/EEC
Technical standards: ..................
EN 60601-1:1998; EN 60601-1-1:2000;
...................................................
EN 60601-1-2:2001+A1:2006;
...................................................
IEC 60825-1:1993+A1:1997+A2:2001;
...................................................
EN 60601-1-4:1997+A1:1999
ENVIRONMENTAL SPECIFICATIONS
Storage: ......................................
temp. range between -10°C and 60°C
...................................................
humidity 0-100% (cond.included)
...................................................
atm. pressure 500 to 1060 hPa
Operating: ...................................
temp. range between 10°C and +40°C
...................................................
humidity 30-75% (non condensing)
...................................................
atm. pressure 700 to 1060 hPa
ELECTRICAL SPECIFICATIONS
KERATRON
™
ONDA
Input voltage: ..............................
between 100 and 240 Volts AC
Frequency: ..................................
50/60 Hz.
Power consumption ....................
70 VA
Fuses: .........................................
2 x T 3.15 AH ( 5 x 20 mm)
TOPOGRAPHY PRECISION
Simulated K-Readings: ...............
Within ± 0.25D on a normal cornea.
Distance error .............................
BFS deviation: with Slit Lamp Adapter typical within
±0.15D
Misalignement............................. Mean deviation on the map: ± 0.1 with misalignement up
to 1 mm °C