Keratron
™
Onda Installation and operating manual
OPTIKON 2000
Cod. 161401EN
9-17
2015-03-27
Rev.C
After the installation check the proper working of software running and test Scout
program. If any error messages appears (Automation error, DLL xxx.dll not found,
etc.)”, the suggested actions are:
Unload the Scout software (see section §9.2).
Reinstall it overwriting the files when prompted.
9.11.2 THE FILE SCOUT.INI
This files co
ntain all the settings you make using the “Option Panel”, and other
settings provided by the SW.
This file must be used if:
1. The SW hangs when starts.
When this occurs, try to rename “Scout.ini” (
do not delete it!
) and restart the
software.
2. Runtime error exporting tests.
When this occurs open the Scout.ini file and then, in the “[Export]” section, check
the “path” key, probably it is pointing to a network path not available at the
moment, or to an external drive not connected. Delete this row
“path=x:\nnnn\mmmm\” and retry.
9.11.3 SHARING THE KERATRON
™
SCOUT AMONG SEVERAL OPERATORS
If more than one user performs tests with Keratron
™
Scout, how can I separate tests
to avoid confusion?
There are two ways to proceed:
Create your own “group” (e.g. your name) into the database, and save every
test you perform in that “group” (
not recommended
procedure
)
or
Create more databases (
recommended
procedure
)
Create a folder for each operator using the operating system function to
navigate within the system (Explore Resources or Manage Resources)
Create one database for each folder (using “Keratron Scout” function, “File” ->
“New Database”).