Keratron
™
Onda Installation and operating manual
OPTIKON 2000
Cod. 161401EN
7-6
2015-03-27
Rev.C
During all of the steps it is always possible to press the [Abort] button (4) and
return to the initial screen shown in Figure 7-1.
Figure 7-11: Screen during calibration
After you have acquired the image, only for the first of the six steps, the operator is
requested to confirm whether the acquired sphere appears clean and free of dust
grains (refer to Figure 7-12).
Figure 7-12: Check that the calibration sphere is clean
If the operator confirms that it is clean, pressing [Yes] will cause the software to
trace the 28 mires, and the screen will appear as shown in Figure 7-13, where the
operator must verify that the reconstruction of the rings, at the edge of the mires, is
complete and lacks any exclusions.
Figure 7-13: Reconstruction of the rings