HiScan Touch
Installation and Operation Manual
OPTIKON 2000
Code 14101XEN
11-15
2014-09-02 Rev.B
Figure 11-23- 3D image with an excessively high noise threshold
The image is too dark. To correct it, it is necessary to reduce the noise threshold. The
recommended value is 30 (or a lower one).