DUKE EASY S.T.01/3AW Rev.0 01/2017
7
-
WARNING: For special operations that require chair movements to be locked, make sure to enable
the specific function (see par. LOCKING CHAIR MOVEMENTS).
-
WARNING: The instructions found in this manual and in the wiring diagram must be followed to
connect a suction unit; the suction unit must have EC markings as per directive 93/42/EEC as amended
“Medical Equipment” and meet international safety regulations IEC EN 60601-1 (Medical electrical
equipment – General safety regulations), IEC EN 60601-1-2 (Medical electrical equipment – Collateral
standard: Electromagnetic compatibility).
-
DANGER: when moving the chair, both in manual and especially in automatic mode, do not place hands
and
feet
near
the
bottom
of
the
chair
to
avoid
crushing
hazards
-
WARNING: during automatic chair movement, the operator must stand near the dental unit.
-
WARNING: When moving the instrument table, assistant table and operating lamp, be careful of the
patient and personnel in the surgery to avoid injuring or crushing the patient or personnel in the
surgery.
-
WARNING: Never sit on the tip or foot of the chair.
-
WARNING: During instrument table movement, patients and personnel in the surgery could be injured
by sharp instrument tips.
-
WARNING: before putting equipment into operation after periods of disuse, accurately disinfect the
air and water supply lines, clean and drain the lines and then proceed with disinfection.
-
WARNING: In the event of prolonged dental work on senior citizens, be careful of the formation of
bedsores.
-
WARNING: When moving the chair and, especially, the head rest, the patient’s hair could become
entangled.
-
WARNING: The equipment is equipped with a general water tap. During use, the tap should be open.
At the end of the work day, the tap should be closed or, alternatively, close the unit’s general supply
tap in the surgery.
DEFINITION OF INTENDED USE
Dental unit intended to treat dental pathologies, intended for professional operator use (dentists).
ADMITTED ENVIRONMENTAL CONDITIONS
ADMITTED ENVIRONMENTAL CONDITIONS FOR TRANSPORTATION AND
STORAGE
The packaged device can be exposed for a period of no more than 15 weeks to the following environmental
conditions:
Temperature range -40 to +70°C
Relative humidity range: 10 to 100% non condensing
Atmospheric pressure from 500 to 1060 hPa