DUKE EASY S.T.01/3AW Rev.0 01/2017
39
Manufacturer’s guide and statement – electromagnetic emissions
The device is intended to operating in the electromagnetic environment indicated below. The device customer or user must
guarantee that it is used in this environment.
Emission test
Conformity
Electromagnetic environment - guide
RF emissions
CISPR 11
Group 1
The device only uses RF energy for its internal operations.
Consequently, its RF emissions are very low and most likely do not cause
any interferences with nearby electronic equipment.
RF emissions
CISPR 11
Class B
The device is suited for use in all environments, including home ones
and those directly connected to a low voltage public mains that powers
buildings used for residential purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuation
emissions/flicker
IEC 61000-3-3
Compliant
Manufacturer’s guide and statement – electromagnetic immunity
The device is intended to operating in the electromagnetic environment indicated below. The device customer or user must
guarantee that it is used in this environment.
Immunity test
IEC 60601 test level
Level of
conformity
Electromagnetic environment - guide
Electrostatic discharge
(ESD)
IEC 61000-4-2
on contact ± 6 kV
in air ± 8 kV
± 6 kV
± 8 kV
Floors must be made of wood, cement or ceramic
tile. If floors are lined with synthetic material,
relative humidity must be at least 30%.
Transistors/Burst
IEC 61000-4-4
± 2 kV for power lines
± 1 kV for input/output lines
± 2 kV
± 1 kV
The mains voltage quality must be that typical of
a retail or hospital environment.
Surge
IEC 61000-4-5
± 1 kV between phases
± 2 kV between phase and
grounding
± 1 kV
± 2 kV
The mains voltage quality must be that typical of
a retail or hospital environment.
Brownouts,
short
blackouts and voltage
variations on input power
lines
IEC 61000-4-11
<5 % U
T
(>95 % brownout in U
T
for 0.5
cycles)
40 % U
T
(60 % brownout in U
T
for 5
cycles)
70 % U
T
(30 % brownout in U
T
for 20
cycles)
<5 % U
T
(>95 % brownout in U
T
for 5 S)
<5 % U
T
0.5
cycles
(10mS)
40 % U
T
5 cycles (100mS)
70 % U
T
20
cycles
(500mS)
<5 % U
T
5S
The mains voltage quality must be that typical of
a retail or hospital environment. If the device
user requires continuous operations during a
blackout, the device should be powered by a UPS
or batteries.
Magnetic field at mains
frequency (50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
Magnetic fields at mains frequency should have
levels typical of a retail or hospital environment.
N.B.: U
T
is the AC mains voltage before the application of the test level
Manufacturer’s guide and statement – electromagnetic immunity
The device is intended to operating in the electromagnetic environment indicated below. The device customer or user must
guarantee that it is used in this environment.
Immunity test
IEC 60601 test level
Level of
conformity
Electromagnetic environment - guide
Conducted RF
IEC 61000-4-6
3 Veff
from 150 kHz to 80 MHz
3 V
Portable and mobile RF communication devices should
not be used near any part of the device including
wires, from the recommended separation distance
calculated with the equation applicable to the
transmitter frequency.
Recommended separation distance:
d = 1,17 √P
d = 1,17 √P from 80 MHz to 800 MHz