Setting Output Conditions
ZX-GT User’s Manual
3
FUNCTION SETTINGS
67
Hysteresis
Set the hysteresis width (difference between operation point and return point) for the
upper and lower limits of the judgments if the HIGH, PASS, or LOW judgment is unstable
near the threshold values.
1
Select the hysteresis setting.
2
Change the current value.
Setting value
Description
00.000 to 59.999 (mm)
Sets the hysteresis width. (default value: 00.100)
Select [HYS].
Confirm the selection.
The sub-display flashes.
Move from one digit to another.
Change the current value.
Confirm the selection.
HIGH
ON
OFF
ON
OFF
ON
OFF
LOW
PASS
HIGH threshold
Measured value
LOW threshold
Output
Hysteresis
Operation point
Return point
SUB
zx_gt.book Page 67 Wednesday, May 30, 2007 3:29 PM