Controller
Specifications
Item
ZX-GTC11
ZX-GTC41
Output type
NPN type
PNP type
Compatible receiver
ZX-GT28R
Compatible emitter
ZX-GT28E11
ZX-GT28E41
Measurement cycle
(*1)
1.5ms (Standard mode (NORM))
0.5ms (High-speed mode (FAST))
(*2)
Number of samples to
average
1/2/4/8/16/32/64/128/256/512/1024/2048/4096
Analog output
(*3)
For current output: 4 to 20mA/F.S., max. load resistance 300
Ω
For voltage output: ±4V, (±5 V, 1 to 5 V
(*4)
), output impedance 100
Ω
Timing input, bank switching
input, zero reset input, reset
input
ON: short-circuited with 0V or 1.5V max.
OFF: Open (leakage current: 0.1 mA max.)
ON: short-circuited with power supply voltage
or power supply voltage -1.5V max.
OFF: Open (leakage current: 0.1 mA max.)
HIGH/PASS/LOW
Judgment output
(*5)
Sync output
(*6)
NPN open-collector output
30 VDC 50 mA max.
Residual voltage 1.2 V max.
PNP open-collector output
30 VDC 50 mA max.
Residual voltage 2V max.
Indicator
Judgment output indicator: HIGH (orange), PASS (green), LOW (orange)
Main display (red)
Sub-display (yellow)
Bank 1/2 (orange), zero reset (green)
Main
functions
Number of
registered setups
2 banks
Measurement
Mode
Interrupted beam width measurement, incident beam width measurement,
outer diameter measurement, center position measurement, IC lead pitch
judgment, IC lead width judgment, specified edge measurement, wire position
measurement, glass edge measurement
Display during
measurement
Measured value, resolution, threshold, voltage output value, current output
value (number of display digits can be changed)
Zero reset
functions
Offset setting of zero reset value, zero reset value memory
Hold
Sample hold, peak hold, bottom hold, peak-to-peak hold, average hold, delay
hold
Timer functions
ON delay, OFF delay, one-shot
Adjustment
functions
Optical axis adjust mode/light intensity writing mode, variable binary level,
variable edge filter, analog output scaling
Calculation
2Possible on up to two Controllers (Calculation Unit ZX-CAL2 is required for
connecting Controllers to each other.)
A-B, A+B, width
Other
Measurement cycle setting, threshold setting, hysteresis setting, initialization,
key lock
Temperature characteristic
±0.005%F.S./°C
Current consumption
150 mA max. (including receiver)
Power supply voltage
24 VDC +10%, -15% ripple (p-p) 10% max.
Dielectric strength
1,000 VAC, 50/60 Hz for 1 min
Insulation resistance
20 M
Ω
(at 500 VDC megger)
zx_gt.book Page 133 Wednesday, May 30, 2007 3:29 PM
Содержание ZX-GT -
Страница 1: ...Smart Sensor Cat No Z263 E1 01 ZX GT Series USER S MANUAL...
Страница 12: ...ZX GT User s Manual 10 MEMO...
Страница 17: ...CONTENTS ZX GT User s Manual 15 MEMO...
Страница 20: ...Operation Step Guide ZX GT User s Manual 18 MEMO...
Страница 42: ...Initializing Controller Settings ZX GT User s Manual 40 MEMO...
Страница 50: ...Functions and Operations during Operation RUN Mode 48 ZX GT User s Manual MEMO...
Страница 74: ...Setting Output Conditions ZX GT User s Manual 72 5 Check that setting of correction values is completed SUB...
Страница 90: ...Special Functions ZX GT User s Manual 88 MEMO...
Страница 157: ...Compliance with EC Directives ZX GT User s Manual 5 APPENDICES 155 MEMO...
Страница 162: ...Quick Reference for Displays ZX GT User s Manual 160 MEMO...