OMICRON
37
Technical data
4.10
Electrical insulation groups
The following chapter shows how the inputs and outputs of
CMC
test sets are insulated against PE and
each other.
B = Basic insulation
R = Reinforced insulation
F = Functional insulation
Insulation designed for pollution degree 2.
BINARY OUTPUT 1 (Relay)
BINARY OUTPUT 2 (Relay)
BINARY OUTPUT 3 (Relay)
BINARY OUTPUT 4 (Relay)
2 x Ethernet
2 x USB
LL out
Ext. Interf.
Mains
100 ... 240 V/10 A
Overvoltage
category II
Output group
CURRENT OUTPUT
VOLTAGE OUTPUT
AUX DC OUTPUT
Binary I/O group
BINARY INPUT 1/2
BINARY INPUT 3/4
BINARY INPUT 5/6
PE
B
F
R
R
R
R
B
F
B
B
B
F
F
F
R
R
F
F
F
F
F
F
F
R
R
R
R
R
R
R
B
B
B
B
B
B
B
B
R
F
F
Содержание CMC 310