CMC 310 User Manual
32
OMICRON
4.5.2
Counter inputs 100 kHz (low level)
The SELV interface connector
ext. Interf.
holds 2 high-frequency counter inputs for up to 100 kHz which
are used for testing energy meters.
In addition, 4 additional transistor binary outputs (
BINARY OUTPUT 11–14
) are available. They are
described in section 4.4.4 "Low-level binary outputs (ext. Interf.)" on page 26.
Pin assignment of the external interface
ext. Interf.
(upper 16-pole LEMO socket); view onto the
connector from the cable wiring side:
2 counter inputs
Maximum counter frequency
100 kHz
Pulse width
>3 µs (high and low signal)
Switch threshold
Pos. edge
Neg. edge
Max. 8 V
Min. 4 V
Hysteresis
Typ. 2 V
Rise and fall times
<1 ms
Maximum input voltage
±30 V
Connection
Socket
ext. Interf.
(rear side of
CMC 310
)
Insulation
Reinforced insulation to all other potential groups of the test
equipment. GND is connected to protective earth (PE).
Pin
Function
Pin 1
Counter input 1
Pin 2
Counter input 2
Pin 3
Reserved
Pin 4
Neutral (N) connected to GND
Pin 5
Binary output 11
Pin 6
Binary output 12
Pin 7
Binary output 13
Pin 8
Binary output 14
Pin 9–16
Reserved
Housing
Screen connection