3.3.5
Status LED A and B
The status LEDs A and B above the
Associate
button are of interest in case of troubleshooting.
A
: yellow status LED
• ON indicates that the test set is ready to be controlled by a computer. The hardware checks in the
test set are finished.
• OFF indicates that the test set is waiting for a software image download. This is the case when the
!
button is pressed while powering up the
CMC
test set.
B
: green status LED
• LED B blinks slowly: The
CMC
test set waits for the TFTP download (
t
rivial
f
ile
t
ransfer
p
rotocol) of
a software image.
• LED B is ON: The TFTP download of the software image is in progress.
• LED B blinks quickly: The computer is writing, for example, the software to the flash memory of the
CMC
test set.
► Do not turn off the
CMC
test set as long as the writing is in progress.
3.3.6
SELV interfaces
All inputs and outputs of the SELV group (SELV =
s
afety
e
xtra
l
ow
v
oltage, defined in IEC 60950-1)
reference to a common neutral (
N
) that is internally connected to the protective earth (GND) of the
housing.
Note:
The hardware for
LL out 1–6
and
ETH2
exists but is not used in the
CMC 310
.
External interface (ext. Interf.):
The SELV interface connector
ext. Interf.
holds 4 additional transistor binary outputs
(
BINARY OUTPUT
11–14). Unlike regular relay outputs, the outputs of
BINARY
OUTPUT
11–14 are bounce-free binary outputs and have a minimal reaction time.
In addition, 2 high frequency counter inputs for up to 100 kHz are available for the testing
of energy meters.
LL out 1–6 (low-level outputs 1–6)
The hardware for
LL out 1–6
exists but is not used in the
CMC 310
.
ext. Interf.