Appendix E. Analog Output Type and Data Format Table
Type Code
Input Range
Data Format
+F.S
-F.S
30
0 ~ +20 mV
Engineering Unit
+20000
0
2’s comp HEX
FFFF
0000
31
4 ~ +20 mV
Engineering Unit
+20000
4000
2’s comp HEX
FFFF
0000
32
0 ~ +10 V
Engineering Unit
+10000
0
2’s comp HEX
7FFF
0000
33
-10 ~ +10 V
Engineering Unit
+10000
-10000
2’s c
omp HEX
7FFF
8000
34
0 ~ +5 V
Engineering Unit
+5000
0
2’s comp HEX
7FFF
0000
35
-5 ~ +5 V
Engineering Unit
+5000
-5000
2’s comp HEX
7FFF
8000