The display will read “Production date dd mm rrrr”. This information cannot be edited. To return to the
Production data menu, depress any push-button, for example . The menu will reappear on the display.
Menu:
Production data, depress / Serial number, depress
The display will read “Serial number xxxxxrr”. This information cannot be edited. To return to the Production
data menu, depress any push-button, for example . The menu will reappear on the display.
Menu:
Production data, depress / Software, depress
The display will read “Software v.xxxxx/xx”. This information cannot be edited. To return to the Production
data menu, depress any push-button, for example . The menu will reappear on the display.
Menu:
Production data, depress / Sensor constants, depress
Constant 1
Constant 2
To page through the menu items, use push-buttons (direction downwards) and (upwards). Select the
desired item and depress . The constant value will appear on the display. Should you wish to change this
value, use push-button to place the cursor under the selected figure and increase/decrease the same
using push-buttons and , respectively. The sensor constants shall be set at the values given on the
sensor rating plate. Unless this not the case, the flow meter calibration is incorrect.
Menu:
Production data, depress / Sensor constants, depress / Constant 1, depress
The display will read “Sensor constant 1 xxxxxxxxx“. The value was set during the meter calibration at the
manufacturing plant and it should not be changed. Depress . The display will read “Value entered Press
any key”. Press any key, preferably . This command will cause the menu Constant 1 / Constant 2 to
reappear on the display. The same procedure applies to setting the value of Constant 2. To return to the
Production data menu, depress push-button .
Menu:
Production data, depress / Excitation frequency, depress
1 – 6.25 Hz
2 – 3.125 Hz
3 – 1.0 Hz
4 – 0.5 Hz
To page through the menu items, use push-buttons (direction downwards) and (upwards). Select the
desired frequency and depress .
Menu:
Production data, depress / Excitation freq., depress / 2 – 3.125 Hz, depress
The Manufacturing data menu will reappear on the display with the item “Excitation frequency” on the first
line.
Menu:
Production data, depress / Suppressed flow, depress
The display will read “Suppressed flow xxxxx“. This parameter is usually set at 0.5% Qmax and may be
increased in cases where it can be demonstrated that while no fluid flows through the meter sensor, the
meter indicates a non-zero flow rate. Using push-button move the cursor to the desired position and
increase/decrease the selected figure by push-buttons and , respectively. Replace xxxx by a number
equal to the desired value of suppressed flow rate. Depress . The display will read “Value entered Press
any key”. Press any key, preferably . The Production data menu will reappear on the display with the item
“Suppressed flow rate” selected.
Menu:
Production data, depress / Language, depress