Design, Assembly and Service Manual
Page 24 of 44
Induction flow meter FMG600 series
The two-line display will always show two of the following basic menu options:
Displayed data
Samples
Analog output
Output function
Electrode clean
Serial line
Production data
Dose setting
Zero setting
100 percent
Exit
Display reading: the basic menu
Use push-buttons and to move upwards and downwards in the menu. As in any meter menu, the
selected item is on the first display line with the initial character blinking.
Depress push-button to enter a subordinated menu, or to edit an item. When in a subordinated menu,
depress push-button to return to the higher-level menu (the “Escape” function). When in the basic menu,
the Escape command will bring forth the possibility to terminate the programming mode via the selection
of the “Exit” item of the basic menu.
7.1. Programming of the basic menu items
7.1.1. Displayed data
The “Displayed data” menu allows the operator to choose which parameters are to be displayed. Using
push-buttons
and , select parameters from the following list. The Flow Rate display cannot be
cancelled.
Menu:
Displayed data, depress
The two-line display unit will always show two of the items from the following menu. Use push-buttons and
to page upwards and downwards through the menu items.
Flow rate
Total
Total volume -
Total difference
Operational time
Percent. Flow rate
Last error
Dosing mode
Menu:
Displayed data, depress / Total volume depress
Display line 1 will read “Do not display”, line 2 “l/s …. l“. Use push-buttons and to go up and down in the
menu items. If you choose “Do not display”, depress to return to the “Displayed data” menu. If you wish to
display If you wish to display “Total ”, select the”Total ” item on the menu; depress push-
button skip line 1,(Do not display) and select line 2, “l/s …. l“ (flow-rate unit … volume unit). Using push-
buttons (downwards) and (upwards), select the desired flow-rate and volume units and confirm the
selection by depressing . The display line 1 will then show “0” and line 2 “0.0“.Using push-buttons
(downwards) and (upwards), select the desired number of decimal positions, confirm the selection by
depressing and return to the “Displayed data” menu.