3.7 Testing/Calibrating Temperature Probes 

Handle hot probes carefully. Use protection for your hands and the surface you will be 

placing the hot probes on after removing them from the well.

When calibrating probes at different temperature points, start at the lowest 

temperature and work up to the highest temperature. Do not jump up and 

down from a very hot temperature to a relatively cooler temperature. This will 

reduce the time it takes for the probe well to re-stabilize after you change the 

setpoint. When placing probes into the well, make sure the probe tip goes all 

the way down to the bottom of the probe well, the full 4.5”. This will insure the 

degree of highest accuracy possible when taking your reading.
After calibrating each probe, remove it from the well and place it on a 

protected surface to cool. If you have another probe to calibrate, place it into 

the probe well and allow the calibrator a few minutes to re-stabilize.

3.8 Cooling Down Your Calibrator After Use

Do not remove the power cord, main line power or turn the  calibrator off until completing  

the cool-down procedure.

When you have completed working with the calibrator you must cool the unit 

down to ambient temperature if you intend to move your unit. 

Cool-Down Procedure:

 

1.  Turn the calibrator to the “standby” mode (See section 3.3) 

2.  The cooling fan may continue to run or turn on after a several minutes if 

the  

 

heater assembly inside the unit has not yet fully cooled down 

to ambient  

 

temperature. Do not remove the power cord or the 

main line power until the  

 

fan stops running permanently. 

3.  It is now safe to remove your unit from the power source and/or return to  

 storage.

CL1000 Series Mini hot point

®

 Dry Block Probe Calibrator

3

3-6

CAUTION:

CAUTION:

Содержание CL1000 Series

Страница 1: ...nt Dry Block Probe Calibrator 230 VAC Model Only CAUTION This product is not intended for medical use or use on humans e mail info omega com For latest product manuals www omegamanual info Shop online...

Страница 2: ...RNING These products are not designed for use in and should not be used for human applications Servicing North America U S A Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 US...

Страница 3: ...Reset Switch 3 3 3 3 Rear Panel Connections 3 3 3 3 1 AC Power Input 3 3 3 3 2 Power Switch 3 3 3 4 Controller Parameters Settings 3 4 3 4 1 Changing Controller Parameter Settings 3 4 3 4 2 Changing...

Страница 4: ...anel 3 1 4 Probe Well Styles 3 2 5 Right Side Panel 3 3 6 Rear Panel 3 3 7 Controller Parameter Factory Default Settings 3 5 8 Approximate Probe Well Heat Up Cool Down Transition Times 3 5 9 Calibrato...

Страница 5: ...nect main power cord or main power source when unit is still hot Do not connect and or operate this unit to a non grounded non polarized outlet or power source This unit is intended for indoor use onl...

Страница 6: ...y appropriate device upon verification of compliance 1 5 General Description The CL1000 is a portable rugged benchtop dry block calibration source with a built in precision PID digital controller The...

Страница 7: ...unting Mount the unit in an ambient environment between the specified 32 to 122 F 0 to 50 C on a bench table top or shelf in a horizontal position and operate at least ten inches from any air obstruct...

Страница 8: ...ower Connection and Requirements Electrical connections and wiring should be performed only by suitably trained personnel Connect to a stable earth grounded AC mains supply of the correct voltage for...

Страница 9: ...s 2 4 2 International 230 VAC 50 60 Hz Models On 230VAC models an International style power cord with the proper color code and approvals is provided with stripped wire ends for connection to the prop...

Страница 10: ...CL1000 Series Installation 2 2 4 Notes...

Страница 11: ...is at a safe temperature for power removal or storage After inserting or removing probes into the probe well allow the calibrator time to stabilize before making your measurement Adding or removing p...

Страница 12: ...be a heater or refrigerator As the process temperature approaches the setpoint temperature the hot or cold process will be pulsed to reduce the corrective measures and minimize overshooting The contro...

Страница 13: ...on then press the reset button firmly If this does not correct the problem contact the factory 3 3 Rear Panel Controls and Connections Figure 6 Rear Panel 3 3 1 AC Power Input The customer connects t...

Страница 14: ...oint temperature to advance more rapidly to a desired value The minimum and maximum setpoints are locked out at 0 and 260 C 32 and 500 F respectively Do not re program the controller s parameter setti...

Страница 15: ...get plate temperature and go to the appropriate column Note that there are X s along the diagonal to ensure the original and target temperatures are the same 3 5 LEVEL 1 LEVEL 2 LEVEL 3 TUNE OFF SP1 P...

Страница 16: ...ell and place it on a protected surface to cool If you have another probe to calibrate place it into the probe well and allow the calibrator a few minutes to re stabilize 3 8 Cooling Down Your Calibra...

Страница 17: ...between your benchtop calibrator and PC Figure 10 Wiring Connections from Calibrator to PC 4 3 CN9 SW Communication Software Calibrators with the C2 option come complete with communication software OM...

Страница 18: ...Fan The fan guard should be cleaned as a minimum annually by using a compressed air source 5 3 Fuse Replacement Disconnect all power from source before attempting fuse replacement For continued prote...

Страница 19: ...2 to 122 F Humidity 0 to 80 RH non condensing Internal Control Sensor Platinum RTD 100 00385 Class A Power Requirements Model CL1000A B C D 3 15 Amps 115 VAC 10 50 60 Hz 275W Model CL1000A B C D 230 V...

Страница 20: ...r is set to its factory default settings c An overheat condition may have occurred Refer to section 3 3 d Unit requires service contact our customer service department 3 Display flashing inPt a Unit r...

Страница 21: ...actual value being measured IEC International Electrotechnical Commission NIST National Institute of Standards and Technology PID Proportional Integral Derivative A three mode control action where the...

Страница 22: ...top design Low cost Model CL900A hot point Dry Block Probe Calibrator Temperature Range ambient 22 to 482 C ambient 40 to 900 F Accuracy 1 5 F Control Stability 0 3 F Power 115 or 230 VAC 10 50 60 Hz...

Страница 23: ...by OMEGA if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limit...

Страница 24: ...bine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industri...

Отзывы: