25
MVX10
4. Place the nosepiece mount ³ (with its objective mounts | facing
upward) where the objective has been by aligning the screw holes.
Using the Allen wrench (for M3 screws), clamp the nosepiece mount with
the provided four clamping screws (M3, 6 mm long) ƒ. (Fig. 21)
# As the screw holes may be hidden behind the objective mounts |,
clamp the nosepiece mount while rotating it. (Fig. 21)
# Before installing the focusing assembly on the pillar, adjust the coarse
and fine focus adjustment knobs so that the hand guard … does not
protrude from the lower end of the focusing assembly †. (Fig. 22)
5. Install the focusing assembly on the pillar, and install the observation
tube and coaxial fluorescence illuminator which have been removed
above in the original positions.
6. Attach two objectives onto their respective objective mounts | by
screwing. (Fig. 21)
Hold the objectives and gently rotate the nosepiece until a click position
where the objective to be used is engaged in the light path.
# When transporting the microscope, do not hold it by the revolving nosepiece.
# The parfocal property when the objective is switched is not completely guaranteed.
# The revolving nosepiece cannot be used in combination with the SZ2-FO stage (because the objectives interfere
with the stage).
Fig. 21
ƒ
…
³
|
Fig. 22
…
†
3
Operation
(Fig. 23)
Fig. 23
4
Caution