}When an SZX series transmitted illumination base is used, the microscope can be used in transmitted light brightfield/
darkfield, focal light and simplified transmitted polarized light observations.
To perform these observations, rotate the mirror unit turret on the coaxial fluorescence illuminator to select the light path
without fluorescence mirror unit (No. 4/TBF ).
The following table shows the main specifications of the transmitted illumination base.
TRANSMITTED LIGHT OBSERVATION
Base
Item
Transmitted Illumination Base
SZX-ILLK
High-Class Transmitted
Illumination Base
SZX-ILLB2
Transmitted Brightfield/
Darkfield Illumination Base
SZX-ILLD2
Max. illumination
field diameter
40 mm
63mm: Brightfield
33 mm: Darkfield
Brightfield observation
Available
Darkfield observation
Not Available
Available
Focal light observation
Available
Not Available
Simplified transmit-
ted polarized light
observation
Available
(Note) The SZX-AN analyzer/SZX-PO polarizer are required.
#The MVPLAPO 2XC objectives cannot be used.
Note 1. When the MVPLAPO2XC objective is used and the zoom ratio is 2.5X or less, insufficiency of brightness in the
peripheral section becomes noticeable. To prevent this, place the diffusion plate
*
( 45 mm, made of plastic)
provided with the microscope body below the stage glass of the SZX transmitted illumination base.
When the SZX-ILLB2 is used and the contrast selector lever is set to the “High” position, irregularities may be
noticeable in the image.
*
A protection sheet is attached to the diffusion plate. Peel off the sheet before using the plate.
Note 2. To reduce background noise during fluorescence observation, it is recommended to use the SP-BW-2 stage
plate (B&W) or the SP-FL fluorescence center plate (with metallic plate).
However, if transmitted light observation is also necessary, use the SP-C stage glass or the SP-FL fluorescence
center plate (with glass plate) instead, and place the ND6 filter ( 45 mm), provided with the microscope, in the
position for the diffusion plate described in Note 1) above. If both the diffusion plate and ND6 filter are used, they