8
Confirm that the error range of measurement values before and after measurement
is within plus or minus 3% from the standard value (2 mm or 0.1 inch).
If the measurement values exceed plus or minus 3%, inspect that the stereo optical adapter
is not loose and the lens is not dirty, and check measurement values again.
Also, if the measurement values differ significantly between before and after measurement,
inspect the adapter in the same way, and measure again if necessary.
2mm
0.1inch
Side-view optical adapter
Forward-view optical
adapter
TIP
If the stereo optical adapter is loose, make sure that the O-ring is not detached (see
“Attaching and removing the optical adapter” (page 24)).
78