TIP
●
Increasing the insertion tube looping amount (bending amount) decreases the maximum
angulation angle limit of the angulation section. Keep the insertion tube as straight as
possible to get the best performance from the instrument.
●
The angulation angle tends to be difficult to achieve under low temperatures.
●
The image noise may be noticeable under high ambient temperatures.
●
A yellow temperature indicator appears when the ambient temperature of the distal end
reaches approx. 80 °C. Be careful that this indicates that the upper limit of the ambient
operating temperature is approaching.
4-6 Adjusting the image display
1
Still image (Freeze)
1
While the live image is displayed, press the [FRZ/REC] button.
The observed image is frozen and the freeze icon (
) is displayed in the upper right corner
of the LCD monitor. Do not insert or pull out the insertion tube while the image is frozen.
2
Press the [FRZ/REC] button.
The image will be unfrozen.
TIP
Freezing an image that contains high-speed action may result in a blurred image.
2
Enlarging the image (Zoom)
The display image can be zoomed for an enlarged view of the observation object.
Pushing the [ZOOM] lever upward enlarges the image by up to 5 times.
To restore the image to the original size, push the [ZOOM] lever downward.
TIP
●
Images are zoomed using the “digital zoom” function. Therefore, the image may become
slightly coarse at higher zoom factors. The image may appear grainier when Menu
→
IMAGE MODE
→
DYNAMIC NOISE REDUCTION is set to “ON”.
●
The zoom function is not available during the spot ranging.
38