2-9-3 Operation
This section describes the functions available when the DP21 is connected to a microscope through a control box.
}The power switches should be set to ON in order of; 1) control box; 2) DP21-CB. Otherwise, the control box cannot be
recognized.
}If the revolving nosepiece (U-D7RES or U-D7REA) is not connected to the control box (U-CBS, etc.), the control box cannot
be recognized.
Auto scale switching
When the objective in the revolving nosepiece of the microscope is switched, the displayed scale is automatically
switched according to the newly selected observation magnification.
To use this function, It is necessary to input the magnifications of the objectives in use in the DP21 in advance. This input
can be done by following the Microscope Setting Wizard that is displayed the first time the power is switched ON. If you
have already started up the DP21, the objective magnifications can be input with the following procedure.
Displaying the scale
Set [ON/OFF] @ in the Menu window to [ON] to display the scale.
Setting the scale magnification
1. Select the scale setting button ² in the Menu window to open the scale
setting window.
2. Select the microscope setting button
3
. The Microscope Setting Wizard
starts.
Operation using the hand switch
Adjust the scale value by pressing cursor buttons
, and change the
adjusted position by pressing cursor buttons . Press the SET/OK
button to advance to the next window.
Operation using the mouse
Click on the position to be adjusted and adjust its value by means of
scrolling of a pull-down menu. The value can also be input directly from
the keyboard. Click on [Next] to advance to the next window.
1
²
³