48
Fig. 56
Fig. 57
Fig. 58
Fig. 59
6
Attaching the Halogen Lamp Housing
(Figs. 56 to 59)
Attaching the Halogen Bulb
}The applicable lamp bulb model is the 12V100WHAL-L (PHILIPS 7724) or
the 12V50WHAL-L (LIFE JC).
1. Fully loosen the clamping screw
1
at the top of the lamp housing using
the Allen screwdriver provided with the microscope frame.
2. Remove the lamp housing
2
by lifting it up.
3. Tilt the bulb socket
3
by 90° in the direction of the arrow.
4. While pushing down the bulb clamping lever
4
down, hold the halogen
bulb
5
with gloves or a piece of gauze, insert the bulb pins
6
straight
and fully into the pin sections
7
on the lamp socket.
Then return the lamp clamping lever gently back to the original position
to clamp the bulb.
To prevent reduced bulb life or cracking, do not touch the bulb with
bare hands. If fingerprints are accidentally left on the bulb, wipe the
bulb with a soft cloth.
5. Fit the lamp housing from up and tighten the clamping screw
1
by
applying downward pressure. (Fig. 56)
Caution for Bulb Replacement During or Right after Use
The bulb, lamp housing and areas near these will be extremely hot
during and right after use.
Set the main switch to “ ” (OFF), disconnect the power cord from
the wall outlet, then allow the old bulb and lamp housing to cool
before replacing the bulb with a new of the designated type.
Attaching the Halogen Lamp Housing (Fig. 59)
1. Loosen the bulb socket clamping screw
1
using the Allen screwdriver.
2. Push in the halogen lamp housing
2
with bulb and tighten the clamping
screw
1
.
1
2
3
4
5
6
7
1
2