34
Fig. 45
Fig. 46
2
Attaching the Modules
(Figs. 45 & 46)
Low-Magnification Fluorescence Mirror Units
}Select the suitable mirror units for purpose of observation by referring to
page 35.
}If you want to fabricate optional mirror units, see page 35.
· Mount the mirror units as indicated in the instruction manual of your
reflected light fluorescence system.
Note that mirror units can be mounted only in every other positions.
Objective
1. Screw a UIS objective
2
into the position on the deeper side of the
U-SLRE sliding revolving nosepiece
1
.
2. Screw a XLFluor2X/340 or XLFluor4X/340 low-magnification fluorescence
objective
3
into the position on the shallower side of the U-SLRE.
Sliding Revolving Nosepiece U-SLRE
1. Raise the revolving nosepiece mount fully.
2. Loosen the revolving nosepiece mount screw
4
on the microscope frame
using the Allen screwdriver provided with it.
3. Align the mount dovetail
5
of the sliding revolving nosepiece with the
revolving nosepiece mount dovetail and gently slide the sliding revolving
nosepiece all the way in from the front as shown in the figure.
4. Clamp the revolving nosepiece by tightening the revolving nosepiece
mount screw
4
.
Swinging Condenser U-SC3/U-UCD8
# Attach the U-SC3 with the top lens swung out.
The condenser top lens should be swung out when using the 2X or
4X objective.
1
2
3
4
5
6