17
BX61WI
Fig. 11
Fig. 12
Fig. 13
Fig. 14
4-3 Stage (IX-SVL2)
1
Placing the Specimen
(Fig. 11)
1. Place the specimen on the center of the stage.
}The optional stage center plate (IX-CP50) makes it possible to observe a
wide range of a big petri dish, etc. (Central hole diameter: 50 mm)
2
Moving the Specimen
(Fig. 12)
1. The specimen can be moved by turning the X-axis knob
1
and Y-axis
knob
2
.
The movement strokes are 50 mm (X-axis) x 43 mm (Y-axis).
3
Setting the Grounding
(Fig. 13)
}In case of electrical physiological experiment, etc., the specimen can be
grounded from the stage.
Prepare a grounding wire
1
and M4 screw
2
and attach grounding as
shown in Fig. 13.
# The screw hole may sometimes be stuck by paint, etc. In such a case,
screw in the M4 screw a few times to expose the metallic thread
inside the screw hole and improve the contact before attaching the
grounding wire firmly.
4
Adjusting the X-Axis/Y-Axis Knob Rotation Tension
(Fig. 14)
}The rotation tension of the X-axis and Y-axis knobs can be adjusted inde-
pendently.
1. Loosen the 2 set screws
1
of a knob using the provided Allen wrench,
hold the stage so that it will not move, then turn the knob to adjust the
tension. Turning it in the direction of the arrow increases the tension and
turning in the opposite direction decreases the tension.
2. After adjustment, tighten the set screws firmly.
# If the tension of a knob is too heavy or too light, skipping or returning
of image may occur during the stage movement.
1
2
1
2
1