4/Service
4-8
4-8
6600 0341 000
08/15/99
Replacing the thermal cutout switch
1. Disconnect the power cord.
2. Place the light source in its vertical position. Remove the four cover screws and
washers .
3. Gently slide the cover off, pulling it back away from the bezel, being careful not to
scratch the finished surfaces.
4. Place the light source in its horizontal position. The thermal cutout switch bracket is
toward the front of the light source (between the front bezel and the lamp).
5. Remove the mains wire harness (it connects the power supply to the PCB).
6. Remove the thermal cutout switch holder screw and washer (this can be best done
through the bulb access door). Slip the switch out of its holder and through the
bracket hole grommet.
7. Remove the connectors (P9 and P2, the long connector on the power supply
board) from their circuit board sockets.
8. Install the replacement thermal cutout switch by reversing steps two through seven
and replacing any cable ties which were removed. To mount the switch properly in
the holder, the clip should be located at the very base of the metal portion of the
switch.
9. Perform the electrical safety procedures detailed near the end of this section.
10. Perform the checkout procedure detailed in section 2/Operation before placing the
BiliBlanket Plus High Output in service.
Replacing the optical filter assembly
1. Disconnect the power cord.
2. Place the light source in its vertical position. Remove the four cover screws and
washers.
3. Gently slide the cover off, pulling it back away from the bezel, being careful not to
scratch the finished surfaces.
4. Place the light source in its horizontal position.
5. Remove the screw on the mode selector slide lever.
6. Through the bulb access door, remove the two nuts on the mode select slide lever.
7. Slide out optical filter bracket through the side opening.
8. Install the replacement mounting bracket by reversing steps one through five.
Replace any cable ties which were removed.
9. Perform the electrical safety procedures detailed near the end of this section.
10. Perform the checkout procedure detailed in section 2/Operation before placing the
BiliBlanket Plus High Output in service.
Содержание BiliBlanket Plus
Страница 6: ...iv iv 6600 0341 000 04 25 00 Notes...
Страница 37: ...4 Service 4 15 4 15 6600 0341 000 08 15 99 3 4 9 CI 17 003 004 Figure 4 4 Front view exploded 8 7 5 10 6 2 1 11...
Страница 40: ...4 Service 4 18 4 18 6600 0341 000 08 15 99 Schematics Figure 4 7 Wiring Diagram CI 17 002...
Страница 41: ...4 Service 4 19 4 19 6600 0341 000 08 15 99 CI 02 001 Figure 4 8 Printed Circuit Board Schematic...
Страница 42: ...4 Service 4 20 4 20 6600 0341 000 08 15 99 Notes...
Страница 45: ...Appendix A 3 A 3 6600 0341 000 04 25 00 Notes...
Страница 46: ...Appendix A 4 A 4 6600 0341 000 04 25 00 Notes...