2 3
5.4 Weights and measures sealing
After a weights and measures official has tested and approved the
scale, it must be sealed by installing the security plate with the security
screw.
Replace the bottom cover and secure with the cover screws.
Paper Seal
A paper seal may be placed over one of the screws at the back of the
scale underneath one foot.
Replace the four feet and turn the scale over.
Lead Wire Seal
For regions requiring a lead wire seal, there are 2 cross drilled screws at
the rear of the scale. The screws can be used in the location as sup-
plied or as an alternate, one screw can be used to replace one of the
counter-sunk screws at the front of the scale. (Under the leveling feet on
small size Ranger Scales.)
5.3 Unlocking the menus (Cont.)
SEAL
High Capacity Scale
On high capacity scales, remove the existing seal and six cover
screws from the bottom of the scale. You will have to unscrew the feet
to access the screws.
Position the scale so the bottom is accessible.
NOTE
: Do the next step carefully.
First, press and hold the unlock switch then, briefly press the
ON/ZERO
OFF
switch at the same time, release the unlock switch after the
segment test is displayed. The menus are now unlocked and settings
can be changed.
LOW CAPACITY SCALE
HIGH CAPACITY SCALE
PAPER SEAL
LOCATION
PAPER SEAL
LOCATION
UNLOCK SWITCH
HIGH CAPACITY SCALE
LEAD WIRE SEAL
LOCATION
LEAD WIRE
SEAL