9
2.3 Locating and leveling the scale
2.4 Connecting power
The scale should always be used in an environment which is free from
excessive air currents, corrosives, vibration, and temperature or humidity
extremes. These factors will affect displayed weight readings.
DO NOT
install the scale:
Next to open windows or doors causing drafts or rapid temperature
changes.
Near air conditioning or heat vents.
Near vibrating, rotating or reciprocating equipment.
Near magnetic fields or equipment that generate magnetic fields.
On an unlevel work surface.
In confined areas, allow sufficient space around the instrument for
ease of operation and keep away from radiating heat sources.
Adjust the leveling feet so that the scale is exactly horizontal and the
air bubble in the indicator is centered.
NOTE
: The instrument should be leveled each time its location is
changed.
Check to ensure the voltage printed on the scale label matches your local
line voltage. If this is not the case,
DO NOT
connect the scale to the
power supply, but contact your responsible OHAUS dealer.
Connect the power cord from the scale to a suitable power source. If the
scale is equipped with an optional internal battery, connect the AC cord
supplied to the rear of the scale.
Allow your scale to warm up for at least 30 minutes
to enable it to
adapt itself to the ambient conditions. If the scale has been stored in a
very cold environment before installation, it may require several hours to
stabilize.
The scale now performs a self-test, displays the software version and the
normal weight display then appears.