Defender 6000 Indicators
EN-11
2.3.2 Scale Base without EasyConnect
TM
Connector
For connecting bases (which do not have the EasyConnect
TM
connector) to an i-DT61PW or an i-DT61XWE, a
load cell cable gland kit (P/N 30379716) is available as an accessory.
Removing the pre-installed Load Cell connector and wiring harness.
Before making the connections, remove the pre-installed load cell connector and wiring harness by following
the following steps.
i-DT61PW
1. Remove the 6 Phillips head screws, and open the rear housing by carefully pulling the front housing
forward.
2. Unplug the white load cell connectors from the main housing (two circles).
3. Open the front housing by removing the 12 Phillips head screws.
4. Unplug the white load cell connectors from the main PCBA board.