EN-38
Defender 6000 Indicators
Reset
[
reset
]
Reset the Print menu to factory defaults.
NO
= do not reset.
YES
= reset
Assignment
[
ASSIGN
]
Demand
[
dEmMAN
]
If
Demand
is selected, the sub-menu
Stable Only
will display.
Set the printing criteria.
OFF
= values are printed immediately, regardless of stability.
ON
= values are printed only when the stability criteria is met.
Auto On Stable
[
ON.StAb
]
If
Auto On Stable
is selected, the sub-menu
Mode
will display.
Set the printing mode.
Load
= prints when the displayed load is stable.
LOAD,;2r
= prints when the displayed load and zero reading are stable.
Auto On Accept
[
ON.ACEP
]
If
Auto On Accept
is selected and the weighing mode is
Check
, values will be printed when the weight is
accepted.
ON.ACep
= printing occurs each time the display is within the acceptable range and stability criteria is
met.
Interval
[
IntEr
]
If
Interval
is selected, the sub-menu
Time
will display.
IntEr
= printing occurs at the defined time interval.
The time interval can be set through the numeric keypad
(
i-DT61XWE
)
.
Settings of 1 to 50000 seconds are available. Default is 1.
Printing occurs at the defined time interval.
MT-Continuous
[
mMt.Con
]
If
MT-Continuous
is selected, the print output will be in the
MT-Continuous
format.
mMt.Con
= printing occurs continuously.
Note:
Refer to Appendix A for
MT-Continuous
format.
C.SUmM
Off = disabled
On = enabled
OH-Continuous
[
OH.Con
]
If
OH-Continuous
is selected, the print output will be in the
OH-Continuous
format.
Note:
Refer to Appendix A for
OH-Continuous
format.
OH.Con
= printing occurs continuously.
SICS
[
SICS
]
OFF
= disable MT-SICS command
ON
= enable MT-SICS command
Note:
Refer to Appendix B for
SICS
commands.
End Print
[
End
]
Advance to the next menu or return to the top of the current menu.
Содержание Defender i-DT61PW
Страница 2: ......