Effective 11/2010
50
Oxygen Generator:
The oxygen generators are designed to run 24 hours a day for many years without failure.
The most common failure mode is molecular sieve contamination due to high moisture
content. These units can be re-built at
OGSI
or new units can be installed in the field.
Replacement in the field is possible and takes about 20 minutes per generator.
Oxygen Booster:
The Oxygen Booster is a single-stage, oil-less compressor that should last two (2) to four (4)
years under normal conditions.
Low Oxygen Tank
pressure (less than
60 psi or 4.1 bar
) will
be an indicator that this booster needs to be replaced. Replacement in the field is possible
and takes about 30 minutes per compressor.
Oxygen Compressor:
The high pressure oxygen compressor is a two-stage, oil-less compressor that will require
maintenance at normal intervals.
OGSI
does sell the spare parts for this compressor as a
kit. Please consult
OGSI
for the spares kit for this machine and have the serial number
for the oxygen compressor (or the
MOGS
Unit) available when you call.