Effective 11/2010
20
Oxygen Compressor
The (RIX) oxygen compressor boosts the oxygen pressure from the oxygen
storage tank to
2200 psi (152 bar)
. It is a two-stage compressor with a
self-contained motor, starting capacitor, pressure switch, pressure relief
valve, and check valve. This is a complex machine and comes with its own
Service Manual
. In
AUTO
mode, it will turn on at about
1900 psi (131 bar)
and turn off at
2200 psi (152 bar)
.
Heat Exchanger
The heat exchanger/cooling fan is used to draw air into the unit and to remove
heat from inside the enclosure. It also cools the compressed air before it is
delivered to the air storage tank. The fan operates while the unit runs.
Oxygen Storage Tank
The storage tank serves as a reservoir for the oxygen, prior to entering the
oxygen compressor. The tank is painted green to represent oxygen as opposed to
the grey color of the air tank. The generator will run according to the pressure in
this tank –
ON
at
70 psi (4.8 bar)
,
OFF
at
90 psi (6.2 bar)
. With the oxygen at
this pressure it can be delivered to the high pressure oxygen compressor or be
sent directly out of the machine through the lower auxiliary oxygen port.
Pressure will usually drop to about 30 psi when the oxygen compressor is
running.
Oxygen Booster
The oxygen booster increases the oxygen pressure prior to it being delivered to
the oxygen storage tank.
Fill Whip Chamber #1
This whip allows oxygen to flow to the small cylinder being filled in
Chamber # 1. It has a
CGA 870
Valve at the end of it.
Fill Whip Chamber #2
This whip allows oxygen to flow to the small cylinder being filled in
Chamber # 2. It has a
CGA 870
Valve on the end of it.
Fill Chamber #1
This is the left side blast chamber that the small cylinders can be filled in.
Fill Chamber #2
This is the right side blast chamber that the small cylinders can be filled in.
Содержание MOGS - 100
Страница 2: ......
Страница 6: ......
Страница 50: ...Effective 11 2010 44 MOGS 100 Filling Manifold Design Installation Optional...
Страница 58: ......
Страница 59: ...Appendix...
Страница 60: ......
Страница 67: ...Effective 11 2010 VII Maintenance Log Date Part Reason for Maintenance Authorized Service Technician Signature...