OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
49
13.
Click “Start 8-Point Cal”.
14. Follow the on-screen instructions.
The software will prompt you to turn the cap 45° and then click OK. After
doing this eight times, the software will display a calibration score. If any
errors occurred, follow the on-screen instructions to correct them.
15. Once the calibration is complete, close the software.
16. Return the cap to the zero position and lock it in place by tightening the
locking screws.
17. Be sure to switch the serial cable from the compass port back to the
ORCADA
®
port.