41
Disassembly
Coupling
1.
Remove the baffle. It may be necessary to put a small screwdriver or
allen wrench through the holes in the baffle to loosen it.
2. Remove the sample port plug from the coupling and clean it with a small
brush in the cleaning solution. Inspect the o-ring and replace it if it is
damaged or worn.
3. Remove the coupling from the stand and clean it thoroughly. Use a brush
to clean all holes, crevices, and threads. Set the coupling on its side
rather than on its ends.
Be careful not to scratch the top and bottom surfaces of the
coupling while cleaning, since they are sealing surfaces.
4. Rinse the coupling and use an air jet to dry it.
Pressure Outlet
Baffle
Sample Port
Stainless Steel Ball
Pressure Inlet