25.
Fill a syringe with test fluid. Inject 15 mL of the test fluid into the “Front”
port.
This step is only necessary when testing drilling fluids to prevent solid
particles from damaging the o-rings. If you are preparing the unit for
a calibration, you can fill the rotor up to the o-ring instead of injecting
addition fluid into the cell.
26. Inspect the o-ring (#130-77-5) on the sample port plug for damage.
Replace it if necessary. Screw the sample port plug into place and hand-
tighten.
27.
Check the “Raw Signal” on the “Calibrate Shear Stress” screen. The