External Synchronous Trigger Mode
In the External Synchronous Trigger mode, two
external triggers are required to complete a
data acquisition. The first rising edge starts the
integration period and the second rising edge
stops the integration and starts the next. Thus,
the integration time is the period between the
two external trigger pulses. After the integration
period, the spectrum is retrieved and available
to the user. As in normal mode, no further
spectra are acquired until the original spectrum
is read by the user.