MNL-1024 Rev A
9
Change the Slit
The Flame-NIR+
allows you to change your spectrometer’s slit to match your application requirements.
Be sure to calibrate your
spectrometer after changing the slit.
1.
Find the SMA connector. If a fiber is attached, remove it.
2.
Use the Allen key to remove the 2 the screws attaching the slit to the spectrometer.
3.
Pull the slit out of the spectrometer.
4.
Put the new INTSMA slit connector into the spectrometer, with the key of the connector on the left
side.
5.
Install the 2 screws again. Use the Allen key to tighten the screws carefully (do not over-tighten).
6.
If necessary, connect the fiber again.
Accessories
Ocean Insight provides a range of standard cables and accessories that connect the Flame-NIR+ to our large range of sampling and
light source accessories. Items specifically designed for the Flame-NIR+ are described here; they are not provided with the Flame-
NIR+ spectrometer and must be purchased separately. Visit us at
for a complete list of products available
for all of your spectroscopy needs.
Cables and Connectors
Cables are available to connect your Flame-NIR+ spectrometer to accessories:
•
Accessory cable for light sources and other accessories (DB15 Connector Cable).
•
Accessory cable for HR-type connector (PAK50 Connector Cable).
•
Breakout board (DD4 -BREAKOUT-BOARD) This breaks out the 40 pin DD4 connector to a set of header pins that can be used
to wire each pin, as required.