MNL-1024 Rev A
28
External Hardware Edge Trigger
Mode
In the External Hardware Edge Trigger
mode, a rising edge detected by the
spectrometer from the External Trigger
input starts the integration period specified
through the software interface. After the
integration period, the spectrum is retrieved
and is ready to be read by the user. If
another trigger is sent a new integration
cycle will begin. If a spectrum request is
not received before the integration cycle
has ended, then that data will be deleted
and a new trigger and spectrum request is
required. Only one acquisition will be
performed for each External Trigger pulse,
no matter what the pulse’s duration is.